Operation procedure to turn ON/OFF the A-weighting of frequency weighting
1. Click the current data (measurement in progress data) to activate it and set [Analysis > Frequency weighting > OFF] in this order on the menu.
2. Open the stored power spectrum text data and reproduce it.
Supplement
When the process of the menu [Analysis > Frequency weighting > A-weighting ON] in this order, displays the result which is secondary calculated data of the original measured data. When the data is saved in [.dat] format, the original data is stored. Therefore, when the data saved in [.dat] format is reproduced, the data which is applied the A-weighting for once is displayed.
Upper raw | Data which is measured in the setting of "A-weighting ON (current data)". |
Middle raw | Reproduce the data which is stored in the txt format at upper raw. (The data is measured in the setting of A-weighting ON, so A-weighting is applied double. Therefore, the overall value is different from the value at upper raw. This value is relied on the frequency distribution of the data, so it can be larger or smaller than original data.) |
Lower raw | Reproduce the data which is stored in [.dat] format at upper raw. The overall value is same as the value at upper raw. |
Revised:2006.11.20