How to store or reset the setting condition of the CF-4500, or copy to another unit?

When you upgrade the software of the CF-4500, please backup the panel condition memory and block memory beforehand. Every setting will be initialized when you upgrade the software.


Operating procedure

 

1.Backup the built-in memory to an external USB memory

 

Press the soft keys of [MAIN] > [MEMORY] > [UTILITY] > [COPY FILES] > [COPY FILES TYPES] in this order, and turn both of [DATA] and [PANEL COND] to ON. Return to [COPY FILES] and execute the copy by selecting Yes of the "start copy" in the [INTERNAL⇒USB] screen. (Please refer to the "CF-4500 Memory Operation".)

 

2.Note of the items which are not stored in the panel condition memory

 

The following settings are not stored in the panel condition memory, so note them beforehand, and manually reset them after upgraded.

 

  • Setting items which are not stored in the panel condition memory but stored after the power is turned OFF

Setting items after [MAIN] > [CONDITION]. *Includes the setting of RS-232C.
Setting items after the [MAIN] > [MEMORY]. * NEW is selected as a default for the START TYPE.
Unless the software is upgraded, the setting conditions of after [CONDITION] and [MEMORY] will be stored regardless of the power ON/OFF.

 

  • Setting items which are not stored in the panel condition memory and initialized every time you turn the power ON

Setting items after [MAIN] > [OUTPUT] > [PHONE].
Setting items after [MAIN] > [ANALYSIS] > [CALC].
Setting items after [MAIN] > [TRACKING] > [SCHEDULE].

Setting items after [PHONE], [CALC] and [SCHEDULE] are initialized when the power is turned ON regardless of the upgrade.

*[PHONE] can be enabled when the optional CF-0453 Band-pass Filter & Monitor Function is added, and [SCHEDULE] can be enabled when the optional CF-0451 Tracking Function is added.

The other panel keys of TRIG, AVG START, START COMPARE and SELECT are OFF every time the power is turned ON.

 

3.Panel condition load method

 

  • Soft key setting

Press [MAIN] > [MEMORY] > [PANEL COND] > [READY LOAD] in this order to turn the [READY LOAD] to ON.

  • Panel condition load method

Press [MAIN] > [MEMORY] > [PANEL COND] > [COPY] in this order to set the Source to USB and Target to INT. Press the soft key [ALL ON] once. Then press [COPY EXECUTE] to execute the copy.

 

4.Reset is necessary for the following items (check list)

 

  • LAN setting

[MAIN] > [CONDITION] > [LAN SETTING]

IP ADDRESS 192.168.3.59
SUBNET MASK 255.255.255.0


  • RS-232C setting

[MAIN] > [CONDITION] > [RS232C SETTING]

BPS 9600
CHARACTER BIT 8BIT
PARITY CHECK NONE
STOP BIT 1
FLOW CONTROL NONE
TERMINATOR CR

 


  • SAVE TYPE (Stored data type)

[MAIN] > [MEMORY] > [DATA] > [SAVE TYPE]

TEXT

OFF
BMP OFF


  • PROTECT (Overwrite prevention of the memory)

[MAIN] > [MEMORY] > [PROTECT]

ON/OFF

OFF

 

 

  • START TYPE (Main unit is in starting up state)

[MAIN] > [MEMORY] > [PANEL COND] > [START TYPE]

LAST TIME

 
NEW NEW
SELECT  

*When SELECT is selected, set the start entry again.

 


  • RELAY LOAD (Reproduce the setting of RELAY when the panel condition memory is loaded.)

[MAIN] > [MEMORY] > [PANEL COND] > [RELAY LOAD]

 

ON/OFF

OFF

 

*Red letters; factory default settings

Revised:2015.05.28