The setting examples of the comparator function of the DG-4240 Digital Gauge Counter.

Set the upper limit and lower limit to the UPPER digital switch and LOWER digital switch of the front panel which become criteria of the pass-fail decision.

 

Pass-fail decision is as follows.

Decision criteria Decision Display and output of decision result
LOWER set value ≥Display value LOWER
  • COMPARATOR display LOWER is lit in red.
  • LOWER output relay (No. 10, 11) is turned ON.
  • BCD OUT connector LOWER output (Pin 21) is turned ON.
LOWER set value < Display value < UPPER set value OK
  • COMPARATOR display OK is lit in green.
  • OK output relay (No. 12, 13) is turned ON.
  • BCD OUT connector OK output (Pin 22) is turned ON.
UPPER set value ≤Display value UPPER
  • COMPARATOR display UPPER is lit in red.
  • UPPER output relay (No. 14, 15) is turned ON.
  • BCD OUT connector UPPER output (Pin 23) is turned ON.

Consider which is the counting, positive (+) or negative (-) when the gauge sensor spindle is pushed in, where is the decimal point display position and where is the zero set point before you set the upper and lower values.

 

(Setting example 1) Range from -0.2 to +0.5 mm is determined as OK.

 

The resolution of the GS-251 and GS-1513A is 10µm. -0.20 is displayed when the measurement value is the lower limit -0.2 mm of the passing range. 0.50 is displayed when the measurement value is the upper limit +0.5 mm of the passing range. The value determined as OK does not include the set values specified by the LOWER and UPPER switches. Therefore, when -00021 is set by the LOWER switch and +00051 is set by the UPPER switch, a value in the range of -0.2 to 0.5 mm is determined as OK.

 

 

 

(Setting example 1) Range from +5 to +10 mm is determined as OK.

The resolution of the GS-5011 and GS-1630A is 1 µm. 5.000 is displayed when the measurement value is the lower limit +5 mm of the passing range. 10.000 is displayed when the measurement value is the upper limit +10 mm of the passing range. The value determined as OK does not include the set values specified by the LOWER and UPPER switches. Therefore, when +04999 is set by the LOWER switch and +10001 is set by the UPPER switch, a value in the range of +5 to +10 mm is determined as OK.

 

 

 


Revised:2000/12/08

© Copyright. ONO SOKKI CO., LTD. All Rights Reserved. | Terms of use | Privacy policy