When the Enhance Function of the Search Function at the Cursor Setting page is ON, which has been opened in this order [Display menu > Cursor Set], the value of frequency saved in TEXT format does not become 1/800 of the frequency range. When the Enhance Function is turned on, the frequency value is displayed in the value which is increased the frequency resolution. The displayed data is saved as it is in the TXT format, so the data which is increased the frequency resolution is saved. When the enhance function is turned off, the data becomes 1/800 of the frequency range, so save it in the TXT format by turning off the Enhance Function.
Caution: The data is saved in every 1/800 of the frequency range, when it is saved in [*.dat] format.
Figure 1: Search enhance function |