Trigger by signal input to an external terminal(PULSE/TRIG)

DR-7100 has trigger function that controls start or stop of recording. Two types of trigger signals are provided as follows.

A. Trigger by signal input to an external terminal(PULSE/TRIG)
B. Trigger by signal magnitude recorded to the specified channel

Here, " A. Trigger by signal input to an external terminal(PULSE/TRIG)" is described. (It is said with "external trigger" as follows)

*Preparation for external trigger measurement

1. Set as follows;「MENU」→「INPUT」→「CH_USE」→「EXT」→「TRG」
2. Set as follows;「MENU」→「TRG」→「TRG_MODE」→「ONESHOT」or「REPEAT」

「ONESHOT」…Trigger function is released after the triggering is applied only once.
「REPEAT」…Trigger function is enabled repetitively

Set as follows;「MENU」→「TRG」→「TRG_TYP」→「TTL+」/「TTL-」/「GATE+」/「GATE-」
Press「TRG」on the panel switch and then press REC button for recording to standby the trigger. There are 4 types of "External trigger" as shown bellows.

 

 

1. TTL+

 

The recording starts by the voltage input of TTL level to a PULSE/TRIG terminal at the timing of signal rising.

 

When the recording time is not set, the recording is continued until it is stopped manually whether「TRG_MODE」is「ONESHOT」or「REPEAT」 When only recording time is set, it ignores the signal which corresponds to the trigger signal during recording according to the set value.

 

When 「TRG_MODE」is set to「ONESHOT」, the recording will stop when it comes to the setup time. When 「TRG_MODE」is set to「REPEAT」, it will wait for the next trigger if the save mode is set to「AUTO」.

 

When both the recording time and interval time are set, it follows the behavior according to the each setup (recording time setting and interval setting) whether「TRG_MODE」is set to「ONESHOT」 or 「REPEAT」.

 

 

2.TTL-

The recording starts by the voltage input of TTL level to a PULSE/TRIG terminal at the timing of signal rising. When the recording time is not set, the recording is continued until it is stopped by manual operation whether「TRG_MODE」is「ONESHOT」or「REPEAT」.

When only recording time is set, it ignores the signal which corresponds to the trigger signal during recording according to the set value.

When 「TRG_MODE」is set to「ONESHOT」, the recording will stop when it comes to the setup time. When 「TRG_MODE」is set to「REPEAT」, it will wait for the next trigger if the save mode is set to「AUTO」.

When both the recording time and interval time are set, it follows the behavior according to the each setup (recording time setting and interval setting) whether「TRG_MODE」is set to「ONESHOTO」 or 「REPEAT」.

 

 

 

 

3.GATE+

 

The recording starts by HI level signal input externally to a PULSE/TRIG terminal. The recording is continued while the signal is HI level. (It stops when the signal becomes Low level.) It ignores the setup time of recording and follows the time duration of HI level signal.

「TRG_MODE」is set to 「ONESHOT」, the recording mode will be released after one time of recording. When「TRG_MODE」is set to「REPEAT」and the save mode is set to「AUTO」, it will wait for the next trigger and repeat GATE recording until the recording mode is released by manual operation.

Save mode is set to 「MANUAL」, the recording mode will be released with recording once even if 「TRG_MODE」 is「REPEAT」. To record repeatedly with trigger, press REC button for recording.

GATE signal
HI Level +2 V to 5 V
LOW Level 0 V to +8 V

 

 

4.GATE-

 

The recording starts by Low level signal input externally to a PULSE/TRIG terminal. The recording is continued while the signal is Low level. (It stops when the signal becomes HI level.) It ignores the setup time of recording and follows the time duration of Low level signal.

「TRG_MODE」is set to 「ONESHOT」, the recording mode will be released after one time of recording. When「TRG_MODE」is set to「REPEAT」and the save mode is set to「AUTO」, it will wait for the next trigger and repeat GATE recording until the recording mode is released by manual operation.

Save mode is set to 「MANUAL」, the recording mode will be released with recording once even if 「TRG_MODE」 is「REPEAT」. To record repeatedly with trigger, press REC button for recording.

GATE signal
HI Level: +2 V to 5 V
LOW Level: 0 V to +0.8 V

Revised:2015/06/16



© Copyright. ONO SOKKI CO., LTD. All Rights Reserved. | Terms of use