Sound Calibrator
SC-2500A / SC-2120A

SC-2500A

SC-2120A
About Sound Calibration
What is a sound calibrator?
It is necessary to ensure that the sound lever meter indicates correct sound level for correct measurement.
A sound calibrator that generates predetermined sound pressure is used to check a sound level meter. There are two types of sound calibrator,
piston-phone type and speaker type. They generate the sound with different sound pressure level and different frequency.
They are classified based on the performance into Class 1, Class 2 etc. and the specifications are determined in the standards (JIS C 1515, IEC 60942).
Important points for using calibrator
- The calibration sound pressure level may be different depending on the type of the microphone. Be sure to check the calibration sound pressure level in the instruction manual before performing calibration.
- A Class 1 Sound Calibrator is recommended for a Class 1 (or type 1) Sound Level Meter and microphone.
- Periodical calibration of a sound calibrator is necessary to secure its performance. Calibration of once a year is recommended. Please contact us for more details.
Features
SC-2500A
- Conforms to IEC 60942:2017 Class 1, ANSI S1.40-2006 (R2011) Class 1, JIS C 1515:2020 class 1
- Used for calibration of Class 1 and Class 2 Sound level meter
- Since the SC-2500A uses the sound pressure feedback control method to control fluctuations in sound pressure caused by static pressure, it can generate a stable sound pressure even if the operating environment changes.
- Cost-effective model
SC-2120A
- Conforms to IEC 60942:2003 Class 2, JIS C 1515:2004 class 2
- Dynamic speaker type
- Simple type for quick operation check, cost-effective model
Specification
*1: This product is discontinued.
*2: SC-0313 adapter attached to MI-3140 1/4-inch preamplifier is required.
*3: The value under the standard environment (standard environmental condition:air temperature: 23 °C, static pressure: 101.325 kPa, relative humidity: 50 %)
Related information