Applicable detectors |
FT-0501, FT-0601, FP-0701,
NP series acceleration pickups, LA series sound level meters, etc. |
Calculation method |
1024-point FFT calculation |
Frequency ranges |
500 Hz range (3.75 Hz - 500 Hz)
2 kHz range (15 Hz - 2 kHz)
10 kHz range (75 Hz - 10 kHz) |
Resolution |
0.05 Hz (in 500 Hz range)
0.2 Hz (in 2 kHz range)
1 Hz (in 10 kHz range) |
Display |
Green LED 6 digits, Display
(r/min) = Input Frequency (Hz) * 60 / Number of pulses per rotation (P/R) |
Accuracy |
±2 * resolution |
Measurement time |
500 ms or less |
Range of pulse rate |
0.5 to 199.5P/R in 0.5 P/R
steps |
Detection of rotary direction |
CW, CCW (when used with FT-0501
detector) |
Exponential averaging |
2, 4, 8 or 16 averages |
Analog output |
Voltage: 0 - 10 V (The full scale speed
can be set arbitrarily within the measurement range.) with load resistance
of 1 kilo-ohm or more |
Comparison output |
Upper & lower limits, rotary
directions
Contact capacity: 30 VDC, 0.1 A |
Monitor output |
Common to the analog output
connector, selected with a bit switch |
Interface |
RS-232C (2400, 4800, 9600,
19200 bps) |
Power supply |
100 - 240 VAC (50/60 Hz) |
Operating temperature range |
0 to +40 °C |
Weight |
1.5 kg or less |
Accessories |
Panel mounting fixture, stand,
terminal block connectors (10-pin and 5-pin) |
Options |
AX-5061 RS-232C interface
cable
FT-0101 V/F pulse output
Detectors |