Could you show me how to set the initial values to the FM-2500 Digital Flow Meter when using it with the FP-2140H Fuel Flow Detector?

You need to set initial values of the FP series Fuel Flow Detector as follows before you use the FM series for the first time. You also need to set initial values in the case of using another FP series Fuel Flow Detector.

 

How to set the initial values to the FM-2500 Digital Flow Meter when using it with the FP-2140H Fuel Flow Detector

 

<Notes>

This is the instruction for a standard model. It is not applicable to a special order model.

This procedure shows from the status at the time of shipment.

This is the setting for measuring instantaneous flow rate (volume). For other measurements, please refer to the instruction manual.

Display update time is set at 1s (at the time of shipment). If you would like to change it, please refer to the instruction manual.

1. Set the date and time. (If it is already set, go to (2)) → Setting procedure for the date and time is here.


2.Set the detector model, factor value, and encoder pulse in this order.

 

Operating procedure

1. Press SETUP (MENU) key. The following Menu setup screen appears.

 

data set

>>Select the slot to use.

If you use more than one slot, you need to set for each slot.

Use up or down and select

Slot1: DF0400

then push ENTER, and the following setup menu appears.

 

2. DF-0400 setup menu appears.

Use up or down and select

 Sensor

then push ENTER.

 

3. Sensor setup menu appears.

Use up or down and select

Select

then push ENTER, and the following detector selection list appears.

4. The following detector selection list appears.

>>Select the detector model to use. (The name of selected detector is highlighted.)

5. Specify the detector model to use.

 

(When you use the FP-2140H, select FP214x.)

Use up or down and select FP214X, then push ENTER.

The selection is established.

6. Set the factor value

 

Use up or down and select

Factor,

then push ENTER.

The leftmost figure of the factor value is highlighted.

>> Set the factor value written in the nameplate of detector or in the test certificate.

When you push left or right, the highlighted position moves.

When you push up or down, the highlighted numerical value increases or decreases.

【Example of 4-digit factor value】
When the factor value is 4-digit number, set the leftmost figure as 0: 09824 (setting example)

 

7. Establish the factor value

Push ENTER, then the factor value setting is established.

 

8.Set the encoder pulse

If encoder pulse is displayed (set) 120*10 → Go to 10.

If it is set (displayed) other than 120*10, or if you would like to set "120", set as follows.

Use up or down and select

  Encoder Pulse

Push ENTER, then selection list of the number of pulses appears.

 

9. Select the target encoder pulse from the selection list.

 

lUse up or down and select 120 * 10.

Push ENTER, then selected encoder pulse is established.

When you select 120*10, decimal places of the displayed value will be two. Ex) 0.00
* If you would like to get the displayed value to two decimal places, select 120.

 

 

Information

 

Instantaneous flow rate, sectional total average flow rate:

 

Display unit : kg/h, L/h
Number of display digits : Max. 7 (Decimal place depends on the flow rate.)

 

Flow detector Display range
FP-2140H/2240HA 0.00 to 9999999

 

Sectional total flow rate, total flow rate:

 

Display unit : g, mL
Number of display digits : Max. 7

 

Flow detector Display range
FP-2140H/2240HA 0.00 to 9999999

 

However, when the encoder pulse is set as follows,

 

Encoder pulse setting Display range
120 P/R × 10 0.00 to 9999999
120 P/R 0.0 to 9999999

 

 

10. Push MENU and back to the measurement data display screen.

 

Press Sens is used for setting of the rated pressure of a pressure sensor. In this case, you do not need to set the rated pressure because a pressure sensor is not used. (Please leave it as the status at the time of shipment.

Revised:2007/06/18



© Copyright. ONO SOKKI CO., LTD. All Rights Reserved. | Terms of use | Privacy policy