How much does the delay time of AC output from the LA-3000 series become?

Approx. 5 ms.

When you need the signal without delay, please use "Through" output.

*Delay time in single mode

 

5.01 ms (Input – AC-Z OUT)
5.01 ms (Input –AC OUT)

 

 

Delay time using 1/3 octave filter

 

Center frequency (Hz) Sampling frequency (Hz) Input-ACOUT delay (ms)
4 kHz 64 kHz 5.137 ms
2 kHz 32 kHz 5.566 ms
1 kHz 16 kHz 6.396 ms
500 Hz 8 kHz 7.959 ms
250 Hz 4 kHz 11.23 ms
125 Hz 2 kHz 17.969 ms
63 Hz 1 kHz 30.859 ms
31.5 Hz 500 Hz 56.25 ms

 


Delay time using 1/1 octave filter

 

Center frequency (Hz) Sampling frequency (Hz) Input-ACOUT delay (ms)
4 kHz 64 kHz 5.078 ms
1 kHz 32 kHz 5.420 ms
500 Hz 16 kHz 6.113 ms
250 Hz 8 kHz 7.344 ms
125 Hz 4 kHz 9.766 ms
63 Hz 2 kHz 15.234 ms
31.5 Hz 1 kHz 25.000 ms

 

Revised:2009.11.16