The delay time of the AC output of the LA-3000 series.

The delay time of the AC output of the LA-3000 series is approx. 5 ms.
If the signal with no delay time is required, please use the Through output.

 

Delay time in Single mode


5.0 ms (Input – AC-Z OUT)
5.0 ms (Input – AC OUT)

 

Delay time in 1/3 octave filter

Center frequency (Hz) Sampling frequency (Hz) Input – AC OUT delay (ms)
4 kHz 64 kHz 5.1 ms
2 kHz 32 kHz 5.6 ms
1 kHz 16 kHz 6.4 ms
500 Hz 8 kHz 7.9 ms
250 Hz 4 kHz 11.2 ms
125 Hz 2 kHz 17.9 ms
63 Hz 1 kHz 30.8 ms
31.5 Hz 500 Hz 56.3 ms

 

Delay time in 1/1 octave filter

Center frequency (Hz) Sampling frequency (Hz) Input – AC OUT delay (ms)
4 kHz 64 kHz 5.1 ms
1 kHz 32 kHz 5.4 ms
500 Hz 16 kHz 6.1 ms
250 Hz 8 kHz 7.3 ms
125 Hz 4 kHz 9.7 ms
63 Hz 2 kHz 15.2 ms
31.5 Hz 1 kHz 25.0 ms

 

Revised:2017.05.31