The delay time of the AC output of the LA-3000 series is approx. 5 ms.
If the signal with no delay time is required, please use the Through output.
Delay time in Single mode
5.0 ms (Input – AC-Z OUT)
5.0 ms (Input – AC OUT)
Delay time in 1/3 octave filter
Center frequency (Hz) | Sampling frequency (Hz) | Input – AC OUT delay (ms) |
---|---|---|
4 kHz | 64 kHz | 5.1 ms |
2 kHz | 32 kHz | 5.6 ms |
1 kHz | 16 kHz | 6.4 ms |
500 Hz | 8 kHz | 7.9 ms |
250 Hz | 4 kHz | 11.2 ms |
125 Hz | 2 kHz | 17.9 ms |
63 Hz | 1 kHz | 30.8 ms |
31.5 Hz | 500 Hz | 56.3 ms |
Delay time in 1/1 octave filter
Center frequency (Hz) | Sampling frequency (Hz) | Input – AC OUT delay (ms) |
---|---|---|
4 kHz | 64 kHz | 5.1 ms |
1 kHz | 32 kHz | 5.4 ms |
500 Hz | 16 kHz | 6.1 ms |
250 Hz | 8 kHz | 7.3 ms |
125 Hz | 4 kHz | 9.7 ms |
63 Hz | 2 kHz | 15.2 ms |
31.5 Hz | 1 kHz | 25.0 ms |