Ono Sokki Company, Ltd.

    Back

What kind of processing is the tracking setting of "ORDER PEAK" and "BAND WIDTH" ?


"ORDER PEAK" and "BAND WIDTH" are functions to perform the tracking of the order component as correct as possible with the size. These functions are suitable when the order does not match to the setting analysis order.

 

■ ORDER PEAK

ORDER PEAK is the function to select the maximum value from three values such as selected order and both sides of it (both sides of the FFT analysis result.) as a tracking data.

■ BAND WIDTH

BAND WIDTH is the function to select partial overall value of the selected number of lines in the BAND WIDTH which sets the specified order as a center line according to the order to be performed tracking. The function is turned OFF when the setting value is "0". The function is enabled when the setting value is "3".The BAND WIDTH function is enables even the PEAK search is turned OFF. The BAND WIDTH function takes priority when the ORDER PEAK and BAND WIDTH are turned ON.

 

(2008.03.18)