| Applicable
standard |
IEC
60942:2003 Class 1/C
JIS C 1515:2004 Class 1/C |
| Method |
Piston-phone |
|
Applicable microphone |
1/2-inch microphone (MI-1211/1233/1234/1431/1432)
1/4-inch microphone (MI-1531*)
* The
adapter which is provided as standard with the MI-3140
preamplifier for MI-1531 is required in use. |
| Nominal
sound pressure level |
114 dB ref.
20 μPa |
| Generated
sound pressure level |
114 dB ±
0.4 dB (at 23 °C/50 % RH/ atmospheric pressure of
101.325 kPa ) |
| Distortion
rate |
2.5 % or
less |
| Nominal
frequency |
250 Hz |
| Generated
frequency |
250.0 Hz ±
0.4 % |
| Output
stabilizing time |
30 seconds
or less after a microphone is inserted. |
| Operation
mode |
Automatic
stop mode (approx. one minute) or continuous mode |
| Operating
temperature range |
-10 to +50
°C |
| Storage
temperature range |
-20 to +70
°C (except for batteries) |
| Operating
humidity range |
25 to 90 %
RH (with no condensation) |
| Storage
humidity range |
10 to 90 %
RH (except for batteries) |
| Atmospheric
pressure range |
65 to 108
kPa |
| Power
requirement |
Type AA dry
cell battery (R6P or LR6) x 3 pcs. |
| Battery
life |
Approx. 2.5 hours (when using R6P type AA dry cell
batteries at 25 °C) |
| Piston life |
Approx. 80
hours(at 25 °C) |
| Outer
dimensions |
60 (W) × 38
(H) × 200 (D) mm |
| Weight |
Approx.
600g (not including batteries) |
| Accessories |
- Barometer
- Type AA dry cell battery (R6P) x 3 pcs.
- Microphone adapter for 1-2-inch microphone
(SC-0312)
- Storage case
|