Main Unit (DS-5100)

40 kHz system 100 kHz system
Maximum number of input channels 48ch (240 ch when the hardware to hardware connection function used) 4ch
Maximum number of output channels 6ch 2ch
Max. number of external input channels (revolution/trigger) 4ch (10 ch when the hardware to hardware connection function used) 4ch
Maximum number of input/ output units* 8 units 3 units
Hardware connection method Daisy chain None
Maximum hardware to hardware connection 5 Units None
*Main unit and battery unit excluded

General specifications
Operating temperature range -10 to 50 ºC (humidity 20 to 80 %RH, no dew condensation)
Storage temperature range -20 to 60 ºC (humidity 20 to 80 %RH, no dew condensation)
Applicable standards CE marking Low Voltage Directive 2014/ 35/EU standard EN61010-1
EMC Directive 2014/ 30/ EU standard EN61326-1
RoHS Directive 2011/ 65/ EU standard EN IEC 63000




Input Unit

40 kHz
Number of input channels DS-0523: 3ch
DS-0256: 6ch
Input terminal BNC
Input format Single end (each channel always isolated)
Frequency range Max. 40 kHz
Sampling frequency Max. 102.4 kHz
Sensor power supply(CCLD) Provided (disconnection detection function is available)
TEDS function Acceleration sensor of IEEE 1451.4 Ver.0.9 / 1.0 and
microphone supported
Power sensor of IEEE 1451.4 Ver. 1.0 supported
Maximum input voltage 30 Vrms (42.4 Vpeak)
Frequency accuracy ±50 ppm
D/A converter 24 bit ΔΣ type
Dynamic range 130 dB
Harmonic distortion Less than 20 kHz -80 dB
Aliasing -93 dB or less
Amplitude flatness Less than 20 kHz ±0.1 dB
20 kHz or more ±0.2 dB
Full scale accuracy ±0.1 dB (at 1 kHz)
Amplitude linearity ±0.0015 % (in full scale)
Channel-to-channel cross-talk -100 dB or less (at 1 kHz, 0-dBVr range)
Channel-to-channel gain accuracy
(within the same range)
Less than 20 kHz ±0.1 dB
20 kHz or more ±0.2 dB
Channel-to-channel phase accuracy
(within the single unit)
Less than 20 kHz ±0.1°
20 kHz or more ±0.7°
Channel-to-channel phase accuracy
(between multiple units)
Less than 20 kHz ±0.6°
20 kHz or more ±1.2°
Anti-aliasing filter LPF 300 kHz, -3 dB


100 kHz
Number of input channels DS-0532: 2ch
DS-0534: 4ch
Input terminal BNC
Input format Single end (each channel always isolated)
Frequency range Max. 100 kHz
Sampling frequency Max. 256 kHz
Sensor power supply(CCLD) Provided (disconnection detection function is available)
TEDS function Acceleration sensor of IEEE 1451.4 Ver. 0.9/ 1.0 and
microphone supported
Power sensor of IEEE 1451. 4 Ver. 1. 0 supported
Maximum input voltage 30 Vrms (42.4 Vpeak)

* 40 kHz unit and 100 kHz unit of input units cannot be mixed.

Frequency accuracy ±50 ppm
D/A converter 24 bit ΔΣ type
Dynamic range 120 dB
Harmonic distortion Less than 20 kHz -80 dB
20 kHz or more -75 dB
Aliasing -93 dB or less
Amplitude flatness Less than 20 kHz ±0.1 dB
20 kHz or more ±0.2 dB
Full scale accuracy ±0.1 dB (at 1 kHz)
Amplitude linearity ±0.0015 % (in full scale)
Channel-to-channel cross-talk -100 dB or less
(at 1 kHz, 0-dBVr range)
Channel-to-channel gain accuracy
(within the same range)
Less than 20 kHz ±0.1 dB
20 kHz or more ±0.2 dB
Channel-to-channel phase accuracy Less than 20 kHz ±0.1°
20 kHz or more ±0.7°
Anti-aliasing filter LPF 300 kHz, -3 dB
 

External Input Unit

Number of output channels DS-0542: 2ch
DS-0544: 4ch
DS-0543: 2ch
Input terminal BNC
Input format Single end (each channel always isolated)
Maximum input voltage 30 Vrms (42.4 Vpeak)




Signal Output Unit

Number of output channels DS-0543: 1ch
DS-0545: 2ch
Output terminal BNC
Isolation 42.4 Vpk
Between BNC ground and frame ground,
and between each BNC ground
Output voltage amplitude ±1 mV to ±10 V
Offset voltage ±10 V
Output coupling DC
Maximum output current 10 mA
Frequency range 40 kHz : 0 to 40 kHz
100 kHz : 0 to 100 kHz

*"MIX IN" on the DS-0545 2ch Signal Output Unit can be used for the products to be developed.

D/A converter 24 bit ΔΣ type
Signal type Sine wave, swept sine, random (no inter-channel correlation), pseudo random, impulse, octave band noise, pink noise, record data
THD and spurious -75 dB or less
(sine wave 1 kHz, amplitude ±1 V output)
Applicable FFT analysis length 64 to 16384 (power of 2)
Zoom analysis Supported (interlocked with zoom analysis range)
Voltage amplitude accuracy Within ±0.2 dB (1 kHz, 1 Vpk, 1 MΩ load)
Frequency accuracy ± 50 ppm
Digital filter Smoothing filter-
Base band: 10th order ellipse / Zoom: 6th order ellipse
Pink filter Analog system: -3 dB/oct ±1.0 dB
(specified at 20 Hz to 20 kHz)
Burst function Provided (continuous/single, 1 ms to 32 s)
Number of burst: 1 to 32767, in increments of 1 cycles
Taper function Provided (1 ms to 32 s)
Disabled when the burst function is enabled




Battery unit

Battery pack Lithium-ion battery pack
manufactured by RRC, RRC2020 99.6 Wh
Charging operation When AC adapter is connected to DS-5100 main unit
External DC power input voltage DC 10 V to 28 V
Option Dedicated charger manufactured by RRC
External DC power cable 5 m
(alligator cable with fuse)